7 replacement of the lamp, Replacement of the lamp, Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual

Page 48

Advertising
background image

48

Instructions for use V 1.5 – 10/2012

9.

Cleaning, disinfection, maintenance

Turn the instrument off with the automatic
mains fuse.

Pull the mains plug.

Slightly lift the sliding window (1) holding it by
the grip (2) and pull it out to the front.

For the technical specification of the lamp,
please refer to Chapter 4. ‘Technical data‘.

Removal of the lamp

The lamp (3) is mounted behind a glare shield (4)

and therefore not visible.

Touch the lamp for better orientation.

Lightly tilt the fluorescent tube down to the left
and pull it out of the clip (5).

Hold the lamp with both hands and pull it to
the left out of the holder (6).

Installation of the new lamp

Hold the lamp in the correct mounting posi-
tion as shown (Fig. 35) and push it to the right
until it engages in the holder.

Lightly push the fluorescent tube upward to
engage in the clip.

Replace the sliding window.

Reconnect the instrument to the mains power
and turn it on.

1

2

3

4

5

6

Fig. 34

Fig. 35

Fig. 36

9.3.7 Replacement of the lamp

Advertising