4 handwheel assembly, Handwheel assembly, Unpacking and installation – Leica Biosystems CM1850 UV User Manual

Page 16

Advertising
background image

16

Instructions for use V 1.5 – 10/2012

4.4

Handwheel assembly

•

Insert the pin (1) of the handwheel shaft in to
the hole (2) of the handwheel.

•

Mount the spring washer (3) on the screw (4)
as shown in Fig. 16.

•

Tighten the screw (4) with an Allen key (5 mm).

•

Attach the cover disc (5) (self-adhesive).

To dismount, proceed in reverse order.

The handwheel including the fixing components are packed in the
cardboard box for the accessories. The handwheel can be dis-
mounted for transporting (e.g. narrow doors).

Only rotate the handwheel when the refrigeration system is on
and the cryochamber is cold.