Instructions for use V 1.5 – 10/2012

4.1

Site requirements

Do not operate in rooms with explosion hazard! To ensure an ad-
equate cooling capacity, the instrument must be set up with at
least 10 cm distance from walls and furniture.

The place of installation must meet the following requirements:

– no direct sunlight,
– mains power socket at a distance no greater than approximately 1.5 m,
– no drafts (air condition outlets, etc.) directly over the instrument,
– even floor,
– mainly vibration-free floor,
– obstruction-free access to the handwheel,
– room temperature max. 35 °C,
– Air humidity must not exceed 60 %.

High room temperatures and excessive air humidity affect the
cooling capacity of the cryostat.

4.2

Transport to the desired location

4.

Unpacking and installation

The instrument must be transported in an upright position only.
When the instrument is tilted, the compressor oil is displaced.
Do not grip the cabinet at the lid. Grip the cabinet only at the

marked locations (

).

The alignment of the adjustable feet is necessary to ensure an un-
obstructed drain of the quick freeze shelf defrosting water.

Lengthwise
movement

Sideways
movement

Tilt

Fig. 4

15

14

16

Move the instrument to the installation site on its wheels (14). Please
note the areas which are reinforced for transporting and grip the cabi-
net only at these locations (Fig. 4).

The adjustable feet (15) can support the weight of the instrument when
tipping at a slight angle (max. 30 °).

At the installation site unscrew the screws in the adjustable feet (15)
using a fork wrench.
This is necessary to ensure stability.

Align the adjustable feet to level the instrument.