Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual

Page 43

Advertising
background image

43

Leica CM1850 UV – Cryostat

9.

Cleaning, disinfection, maintenance

•

Loosen the screws (3) with an Allen key (4 mm).
Do not loosen the screws (7) for removing the
microtome.

•

Disconnect the coarse feed motor plug (1) by
pulling the metal head.

•

Disconnect the temperature sensor (2) from
the microtome.

1

2

3

4

5

7

8

8

•

Slightly lift the microtome and pull it to the left
to disengage the plastic coupling (5) connect-
ing the axes.

•

Take the microtome out of the cryochamber.

Fig. 25

Fig. 26

Advertising