Daily use of the instrument, 1 specimen freezing, 1 quick freeze shelf – Leica Biosystems CM1850 UV User Manual

Page 26: 2 stationary heat extractor, Specimen freezing, Quick freeze shelf, Stationary heat extractor

Advertising
background image

26

Instructions for use V 1.5 – 10/2012

7.

Daily use of the instrument

7.1

Specimen freezing

Select the appropriate sectioning temperature (cryochamber tempera-
ture) for the sample material (please refer to 7.4 Temperature selection
chart‘).

7.1.1 Quick freeze shelf

The cryochamber has a quick freeze shelf (5, Fig. 17) for up to 10 specimen discs.
The temperature of the quick freeze shelf is always lower than the cryochamber
temperature.

Cut the specimen to size.

Activate the Peltier element (4), if available - it
may take up to 40 seconds until the maximum
refrigeration output is available.

Apply enough cryocompound to a specimen
disc at room temperature.

Place the specimen on the disc and orient.

Place the specimen disc in one of the holes of
the quick freeze shelf and freeze the speci-
men at a low temperature.

Once the specimen is frozen, insert the speci-
men disc in the specimen head
(Fig. 18) and start sectioning.

7.1.2 Stationary heat extractor

Fix the support (1) of the heat extractor (2) by tightening the 2 screws to
the threaded holes on the left sidewall of the cryochamber and insert the
heat extractor.

Lower the heat extractor cylinder on the specimen surface. After ap-
proximately 30 seconds contact, the specimen will be entirely frozen.

Once the specimen is frozen, place the heat extractor on the parking
station (3).

1

2

3

4

5

Fig. 17

Advertising