17

Leica CM1850 UV – Cryostat

5.1

Connection to mains power

During the start-up of the compressor the nominal voltage must not
drop below the values specified in the ‘Technical data‘.
Please note that the compressor requires a start-up current be-
tween 45 and 50 A.
Therefore, the electric circuit at the installation site must be in-
spected by an electrical engineer to ensure that it meets the re-
quirements for a smooth operation of the instrument.
A constant adequate power supply to the instrument must be en-
sured at all times.
Failure to comply with the above will cause severe damage to the
instrument.

The electric circuit at the place of installation must be protected sepa-

rately.

Do not connect any other consumers to this electric circuit.

5.2

Prior to operation

Before connecting the instrument to the mains power, please check if

the local mains voltage complies with the power rating indicated on the
name plate of the instrument.

Place the storage shelves in the cryochamber.

Place the section waste tray and brush shelf in the cryochamber.

Place the knife holder base on the microtome base plate.

Insert the knife holder and clamp it on the base plate For details, please

refer to the separate instruction manual for your knife holder).

Open the knife box with the knife and place it in the cryochamber for

precooling.

Place all tools needed for specimen preparation in the cryochamber.

Close the sliding window.

Connect the mains plug to the mains power outlet at the wall.