Troubleshooting – Leica Biosystems CM1850 UV User Manual
Page 38
38
Instructions for use V 1.5 – 10/2012
No or insufficient refrigeration
Scraping noise at the slot cover
of the microtome
Both disinfection indicator lights
blinking alternately.
- Compressor defective.
- Leak in the cooling system.
- Inappropriate site conditions.
- Ventilation grilles of the con-
denser dirty.
- Friction between the slot cover
and the microtome housing.
- UV radiation provided by UV
tube no longer sufficient for
disinfection.
- Call technical service.
- Call technical service.
- Check site conditions as de-
scribed in section 5.2.
- Clean the ventilation slits as
described in section 10.3.1.
- Apply cryostat oil to the slot
cover and distribute by turning
the handwheel or with a cloth.
- Replace UV tube following the
manufacturer’s instructions.
Problem
Cause
Remedy
8.
Troubleshooting