Instrument components and specifications, 1 technical data, Technical data – Leica Biosystems EG1150 H User Manual

Page 10

Advertising
background image

10

Instructions for Use V 2.5 RevF – 09/2014

3. Instrument Components and Specifications

3.1 Technical data

General data

Approvals:

The instrument-specific approval marks are located on

the rear panel of the instrument next to the type plate.

Voltage of power supply:

100-120 V, 50/60 Hz

230-240 V, 50/60 Hz

Power consumption:

1100 VA

Protection class

1

: I

Pollution degree

1

: 2

Overvoltage category:

II

Operating temperature range:

+18 °C to + 40 °C

Operating temperatures:

55 °C to 70 °C, adjustable in 5

K increments.

Relative humidity:

Max. 60 % non-condensing.

1)

according to IEC-1010, UL 3101, EN 61010

Fuses

Standby switch:

Circuit breaker manufactured by ETA, model

3120-F421-P7T1-W01D-5A

Fine-wire fuses 6.3 x 32 mm:

2x T2.0 A; 2x T4.0 A; 4x T5.0 A; 2x T6.25 A

Fuse type:

Schurter: Type Fst

Dimensions and weights

Dimensions:

Height:

360 mm

Width:

500 mm

Depth:

640 mm

Weight:

Approx. 22 kg

Capacities

Paraffin reservoir:

Approx. 3 l

Cassette tray:

Approx. 100 cassettes

Embedding mold tray:

Approx. 50 embedding molds

Programmable parameters

Temperature:

Paraffin reservoir/dispenser

(heating-up time = 4 h)

Embedding mold tray, cassette tray, work top

Time:

Working day, current weekday

Working times (start, end), current time