5.1 Instrument parts/functions

Paraffin reservoir (1)
The paraffin reservoir has a capacity of approx.

3 liters. The paraffin temperature can be set be-

tween 55 °C and 70 °C in 5 K increments. The lid

should always be in place otherwise the adjusted

temperature cannot be maintained.
An overtemperature cutout is provided to prevent

overheating the paraffin if temperature control

fails.

A built-in filtering screen prevents any particles

contained in the paraffin from getting in the par-

affin block.

5. Operation

Fig. 10

1

2

12

13

11

14

13

1

2

Dispenser (2)
The dispenser is heated separately. The tem-

perature setting of the dispenser and paraffin

reservoir is coupled.

The quantity of paraffin released from the filler

tube (14) can be adjusted continuously with the

metering screw (11).

The dispenser handle (12) is used for manually

operating the paraffin flow. It is provided with

a pressure clip (13). The pressure clip is height-

adjustable and can be shifted to the side (left or

right) depending whether the dispenser handle is

released with the mold or the finger .

The dispenser handle can be operated by simply

pushing the mold (or the finger) against the pres-

sure clip. Slightly pushing the handle backwards

will open the dispenser valve . Once released,

the handle flips back in the original position and

the valve shuts.

The flow cannot be stopped complete-

ly with the metering screw (11). It must

not be turned while cold!

Recycled paraffin may NOT be used in

the Leica EG1150 H due to the danger

of contamination.