Troubleshooting, 1 possible faults, Possible faults – Leica Biosystems EG1150 H User Manual

Page 34

Advertising
background image

34

Instructions for Use V 2.5 RevF – 09/2014

7. Troubleshooting

7.1 Possible faults

This chapter will help you remedy problems that may occur when working

with Leica EG1150 H.
If a problem cannot be remedied by following the instructions in this chap-

ter, please contact your Leica technical service center.
For further instructions, please refer to Chapter 8.
The table below lists the most common problems that may occur as well as

possible causes and remedies.

Problem

Possible cause

Corrective action

1. Display:


The

message appears

in the display.

The colon flashes.

- The instrument is in the heating

phase (standby mode).

- This is not an error!

The instrument will switch

to the operating mode at the

programmed start time.

2. Instrument doesn't work.

- Standby switch is not turned on

or

- Cut-out function of the standby

switch has been activated.

- ON/OFF button has not been

pressed long enough.

- Standby switch ON.

- Check instrument's connec-

tion, then turn standby switch

on.

- Hold down ON/OFF button

for several seconds (at least

2 secs)