7. Troubleshooting
7.1 Possible faults
This chapter will help you remedy problems that may occur when working
with Leica EG1150 H.
If a problem cannot be remedied by following the instructions in this chap-
ter, please contact your Leica technical service center.
For further instructions, please refer to Chapter 8.
The table below lists the most common problems that may occur as well as
possible causes and remedies.
Problem
Possible cause
Corrective action
1. Display:
The
message appears
in the display.
The colon flashes.
- The instrument is in the heating
phase (standby mode).
- This is not an error!
The instrument will switch
to the operating mode at the
programmed start time.
2. Instrument doesn't work.
- Standby switch is not turned on
or
- Cut-out function of the standby
switch has been activated.
- ON/OFF button has not been
pressed long enough.
- Standby switch ON.
- Check instrument's connec-
tion, then turn standby switch
on.
- Hold down ON/OFF button
for several seconds (at least
2 secs)