24

Instructions for Use V 2.5 RevF – 09/2014

Pressing the foot switch opens the

valve, releasing it closes it. This keeps

the operator's hands free to work with

the instrument.

The flow volume can be adjusted with

the metering screw (11).

The dispenser handle (12) is not re-

quired when using the foot switch and

can be folded upwards.

Proceed as follows:
• Set the metering screw (11) to mini-

mum.

• Carefully fold the dispenser handle

(12) back/up with your thumb and

index finger.

5. Operation

Foot switch (22) optional

Fig. 14

12

22

11

12

14

14

The foot switch (22) actuates the mag-

net valve of the dispenser.

It must be connected as described in

Chapter 4.6

.

The plug of the foot switch

MUST be screwed to the

socket. Otherwise hot paraf-

fin can escape, even if the

switch is not activated.

Take care when folding back

the dispenser handle! Hot

paraffin may come out of the

filler tube (14).

DANGER OF BURNS!