On the back there are five dif-

ferent fuse holders (

Fig. 24

).

One spare fuse must be se-

lected from the five spare

fuses available that has the

marking 2A written on it.

(see

Fig. 25,

left)

Only the spare fuse with the

marking 2A may be inserted

in this fuse holder (T2A).
This is also applicable in the

same way for the other fuse

holders.

Fig. 25

Select the correct spare fuse

The metal caps on the spare fuses are labeled

- the labeling looks similar to that shown in the

figure above, but can vary depending on the type

of fuse.
The labeling important for the correct allocation

stands for the current intensity to be protected by

the fuse (in the marked example: T2A)

This marking (2A) on the fuse and the labeling next to the fuse holder on the rear of

the instrument MUST agree!

9

10

Change fuse
Using a screwdriver (8) lightly press in the fuse

holder (9) and then turn anti-clockwise one

1/4 turn and let go.
The fuse holder is pressed out and can be

removed.

Take the defective fuse (10)

out of the fuse holder (9) and

replace it by the selected

spare fuse.
Insert the fuse holder with

the spare fuse, press in

using screwdriver (1) and

secure by turning it clock-

wise 1/4 turn.
Replace fuse cover (6,7).

9

8

Fig. 26