Operation – Leica Biosystems EG1150 H User Manual

Page 22

Advertising
background image

22

Instructions for Use V 2.5 RevF – 09/2014

Paraffin collection tray (8)
Two indirectly-heated paraffin collection trays

for excess paraffin are located under the work

surface.

5. Operation

8

6

7

16

17

Fig. 12

8

When working with half-open lids (17),

raise the temperature one level to ensure

that the paraffin remains molten.

Only operate the cassette and em-

bedding mold tray and paraffin res-

ervoir with lid; otherwise the set

temperature cannot be maintained.

The paraffin collection tray must be

emptied daily. To prevent contami-

nation, do not reuse paraffin collect-

ed in this tray.

If the instrument is operated without

the paraffin collection trays there is

risk of burning.

Embedding mold tray and cassette tray (16)
Depending on the preferred working direc-

tion, the two heated instrument trays (6, 7)

can be used for embedding molds or cas-

settes. The temperature is adjustable between

55 °C and 70 °C.

A removable tray (16) for cassettes or embed-

ding molds can be placed in each instrument

tray.

Each receptacle can hold approx. 100 cas-

settes or approx. 50 molds.

A lid (17) is provided for each of the trays (16)

to prevent loss of heat and contamination in-

side the tray. For easy access, the lid can be

slid back and folded up.