Operation – Leica Biosystems EG1150 H User Manual

Page 21

Advertising
background image

21

Leica EG1150 H – Paraffin Embedding Station

Work top (3)
The temperature of the work top can be adjust-

ed between 55 °C and 70 °C in 5 K increments.

This includes the embedding area, forceps

holder (5) and the cooling spot (4).

The working area has a groove around it and

several drain holes (15), through which excess

paraffin rapidly drains.

Cooling spot (4)
The cooling spot is an integral part of the work

area

.

Its position directly in front of the em-

bedding area enables convenient working in

ergonomically most favorable conditions.

To orientate the samples, the mold is filled

about one third with liquid paraffin. The liquid

paraffin begins to solidify rapidly on the cool-

ing spot.

While the paraffin is semi-liquid, the sample

can be oriented as required. Finally, the mold

can rapidly be filled up with paraffin.

5. Operation

3

3

4

5

5

4

15

Fig. 11

The forceps holder is constantly heated

to approximately 70 °C.

Risk of burning!

While orienting the tissue, the paraf-

fin should not become too solid, as

this may cause different phases in

the finished block, including fissures

inside, as a result of which the block

may break during sectioning.

Forceps holder (5)
The removable forceps holder under the dis-

penser can accommodate up to 6 forceps.