2 switching the instrument on, Switching the instrument on, Operation – Leica Biosystems EG1150 H User Manual

Page 25

Advertising
background image

25

Leica EG1150 H – Paraffin Embedding Station

The heaters become active and the LEDs indicate the values last set.
The target value LEDs for the heating areas (1, 3, and 4 in

Fig. 18

) flash

quickly for the duration of the heating phase (cf. "Display of the heating

intervals").

The LED for the paraffin reservoir (2 in

Fig. 18

) flashes in 1-second inter-

vals and thereby indicates the intensified heating of the paraffin reser-

voir (for 4 hours).

If temperature settings are modified during the heating phase, the in-

strument will store the new settings and the heating phase will then

correspondingly take longer.

Fill the paraffin reservoir with paraffin.

5.2 Switching the instrument on

Activate the standby switch (left-hand side, next to control panel)

(see

Fig. 15, 17

).

• All LEDs of the operating panel light up briefly and the installed software

version is shown on the display for around 2 seconds.

All displays go out, the unit changes to standby mode.

The green LED in the ON/OFF button indicates that the instrument is

ready to operate.

5. Operation

Fig. 15

• Press the ON/OFF button down for approx. 2 seconds to

switch over to the operating mode. The time indicator flashes

on the display, indicating that the instrument has been dis-

connected from the power supply.

Press any key to acknowledge.

Fig. 16

Prior to leaving the factory, the Leica EG1150 H is tested thoroughly under laboratory condi-

tions. For this reason, you will find a small quantity of clean, hardened paraffin. You can work

with this paraffin without any problem.

The unit's normal operating mode is STANDBY/ON; i.e. switch on/

off with the ON/OFF key only.

Use the standby switch only if the device is to be switched off for

an extended period.