Troubleshooting – Leica Biosystems EG1150 H User Manual

Page 35

Advertising
background image

35

Leica EG1150 H – Paraffin Embedding Station

7. Troubleshooting

Problem

Possible cause

Corrective action

3. Paraffin reservoir

Paraffin does not melt or melts only

very slowly.

- Selected temperature too low.

- No enhanced mode set.

- Temperature indication incor-

rect, or paraffin reservoir is not

heated.

- Incorrect start time selected.

- Fuse for the heater of the paraf-

fin reservoir activated.

- Technical fault.

- Increase temperature for par-

affin reservoir.

- Activate on-demand heater.

(see

Chapter 5.6

)

- Check temperature settings

and repeat if necessary.

- Check start time.

- Check fuses and replace as

required. (see

Chapter 7.3

)

- Contact customer service.

4. Paraffin dispenser

No paraffin flow.

Paraffin drops when valve is shut.

Non-homogeneous paraffin

flow (e.g. air pockets).

- Paraffin not completely molten.

- The paraffin outlet is clogged or

the magnetic switch defective.

- Air bubbles in dispenser at start

of operation

- Wait until paraffin is entirely

molten; then check again.

- Contact customer service.

- Keep dispenser open at maxi-

mum flow for a period of time.

Advertising