Leica EG1150 H – Paraffin Embedding Station

5. Operation

5.4 Operating modes

Standby mode
The display and all LEDs are switched off in standby mode. The unit switch-

es on or off automatically at the programmed start and ending times.

Only the green LED (21) in the ON/OFF button indicates that the instrument

is ready to operate.

Enabling operating mode
Hold down the ON/OFF button for approx. 2 seconds.
• The instrument switches from standby mode to operating mode. The

LEDs of the currently selected values are illuminated, the display reads

the current time.

21

In 24-hour mode, activate enhanced mode (see

Ch. 5.6

) to acceler-

ate the melting process after replenishing solid paraffin.

Important

The timer is disabled if the end time is before the start time.

For example, start time: 08:00 and end time: 06:00.

24-hour mode – working in shift mode
If the switch-on and switch off times of the timer are set to the same value,

the instrument will run continuously, even on days that are not defined as

work days.

Example: Start time = 00:00 and end time = 00:00.

To set the timer, see

Chapter 5.5

The time in the display flashes when switching to operating mode

if the device was disconnected from AC power for an extended

period (> 5 min).

Acknowledge the flashing display by pressing any button and

check whether all required heating phases (especially the paraffin

reservoir) are complete.