1 u frame, U frame, Incisor bar – Leica Biosystems ANGLE TWO User Manual

Page 10

Advertising
background image

Components

Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

10

3.1 U Frame

Figure 4 shows the Angle Two U frame with a rat adapter and 45° ear bars attached.

Some of the part names in the figures below link to sections of the manual where the part is

discussed in more detail.

Figure 4. The Angle Two U frame

The animal is held in the instrument U frame with the head at the closed end of the ‘U’ and the tail

at the open end. The ear bars, nose clamp and incisor bar attached to the U frame secure the

animal during surgery.

Adapters for a range of species are available (see www.myneurolab.com).

Incisor Bar

The animal’s incisor teeth are placed over the incisor bar. The bar can be adjusted horizontally and

vertically using the thumbscrews indicated below.

Figure 5. The incisor bar can be moved vertically and horizontally

Incisor Bar

U Frame

Nose Clamp

Ear Bars

Incisor Bar

Vertical control

Horizontal

control