5 vernier scales, Vernier scales – Leica Biosystems ANGLE TWO User Manual

Page 22

Advertising
background image

Components

Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

22

3.5 Vernier Scales

All Leica stereotaxic instruments have vernier scales for manual measurements. While the

instrument software makes all calculations for you, some manual measurements might be needed,

for example to center the ear bars side to side.

To use the scales follow the example below.

The main scale, graduated in millimeters, is on the ear bar, and the vernier is on the U frame. To

measure the position of the ear bar:

1. Locate the zero on the vernier scale and read below on the ear bar scale to get the first

measurement – in the example it measures 1.2 cm.

2. Use the vernier scale to measure the second decimal place. To do this locate the line of the

vernier scale that directly aligns with a millimeter line on the main scale below (only one can

align, the rest are slightly off). The third line of the vernier scale aligns with the main scale

beneath, therefore the measurement is 1.23 cm.

Figure 22. Using the vernier scale

First measurement on the

ear bar scale is 1.2 cm

The third line of the vernier scale on the

U frame aligns with the ear bar scale

Advertising