Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

40

15. When the AP tilt is aligned, select ll To AP on the computer screen (see Figure 28).

16. Click Setup and click Zero Angles to zero the Tilt and Rotation.

Figure 45. The set up dialog and Zero Angles button

NOTE: if you cannot vertically align the instrument contact Leica Microsystems.

ML Tilt

13. If the tilt is in the ML direction position the dial indicator spring against the side face of the T-

bar (see Figure 47).

Figure 46. Tilt (1) is in the ML direction (2)