Leica Biosystems ANGLE TWO User Manual
Page 39
Alignment
Leica Angle Two
™
User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010
39
AP Tilt
13. If the tilt is in the AP direction use the ML drive to position the dial indicator spring in the
middle of the T-bar. Do not move the AP drive.
Figure 43. Tilt (1) is in the AP direction (2)
14. Move the spring to the bottom of the bar and then all the way up using the DV drive.
The instrument is vertically aligned if the dial arm only moves several divisions back and forth, but
stays roughly in the same position.
The instrument is
not
vertically aligned if the dial arm continuously moves in one direction.
To select a new vertical position:
(i) Loosen the tilt clamp (see Figure 35) and carefully move the manipulator into a new
position.
(ii) Retighten the clamp and repeat step 14.
Figure 44. Move the dial up the T-bar to check the vertical alignment