Tilt – Leica Biosystems ANGLE TWO User Manual

Page 17

Advertising
background image

Components

Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

17

Tilt

The primary tilt plane can be in either the AP or ML directions (see Figure 15). Always remember to

update the software with the primary tilt direction (“Tilt Direction” on page 29).

To adjust the tilt loosen the tilt clamp and move the manipulator to the desired angle, then

retighten the clamp. A scale below the clamp shows the amount of tilt.

Figure 15. Tilt movement in the AP (left) and ML (right) directions

Changing From AP to ML Tilt

1. Unlock the swing-out clamp and swivel the manipulator 90° so the probe faces the front of the

instrument (see Swing-Out).

Figure 16. The manipulator swung-out to the front of the instrument

Swing-out clamp

Rotation clamp

Tilt encoder