4.2 Modes

Four modes are used during surgery: Set Target, To Target, Atlas and Vertical. Each mode is

used for a specific function and has its own set of coordinate boxes that appear beneath the mode

selection area. Click the appropriate mode to activate it for use.

Figure 25. The four modes and coordinate boxes

Set Target

Select Set Target mode to enter the position of the target site – where the probe will go.

The target site appears as a red crosshair on the coronal plate and its ML, AP and DV coordinates

are displayed in the Target (Red) coordinate boxes (see Figure 25).

Target coordinates can be edited at any time and saved. To save target coordinates click Save

Named Point (see “Saving” on page 31).

There are three ways to enter the coordinates:

1. Select the coronal plate showing the target site and click on a position in the map to set it as

the target.

2. Use saved coordinates – double click on saved coordinates located across the bottom of the

screen. The coordinates are automatically entered and a red crosshair appears (see Figure 26).