Software Overview

Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

27

3. Manually enter the coordinates into the Target (Red) boxes.

Figure 26. Target coordinates can be entered manually into the coordinate boxes, loaded from

saved coordinates or by positioning the mouse on the screen and clicking at the position

If performing bilateral surgery, after the first target site has been located and probe placed, click

+/- beside the ML coordinates box. The target site is moved to the opposite side of the coronal

plate and the To Target coordinates are automatically updated.

To Target

The To Target mode shows the distance that the probe must move in the ML, AP and DV axes to

reach the target site. The probe’s current position is shown as a blue crosshair, and the target site

as a red crosshair. The To Target coordinate values reduce as the manipulator moves closer to the

target site, until they read zero – indicating the target site has been reached.

Figure 27. The distance the probe needs to move to reach the target site is shown in To Target mode

Target

coordinates

Saved

coordinates

Target

position

Probe

position

Distance to

target

Target site