Alignment

Leica Angle Two

User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010

37

Horizontal Alignment

The ML horizontal path of the probe is controlled by the rotation of the instrument. This part of the

alignment procedure checks the ML horizontal path is perpendicular. To test this the dial indicator is

moved horizontally along the length of the T-bar.
10. Use the AP drive to position the dial indicator spring against the T-bar.

Figure 40. The dial indicator spring against the T-bar

11. Continue to move the AP drive so that the dial indicator arm completes one full rotation. Stop

moving the drive when the dial reads zero.

Figure 41. Dial indicator at zero

AP drive

Dial indicator

spring

T-bar