MTS TestLine User Manual
Page 58
Advertising
a) Select Station Setup button > Channels > Ch1 > Displacement > Channel Input Signals Button
> Limits tab.
b) Set the Upper Limit to 25 mm.
Note: You must press Enter after entering the value in the text box.
c) Set the Upper Limit Action to Interlock.
d) Set the Lower Limit to -25 mm.
e) Set the Lower Limit Action to Interlock.
f) Ensure that the actuator will be able to move 50 mm in each direction without contacting any
obstructions. If there are obstructions that could be contacted, move the actuator or set narrower
limits.
2. Set narrow force limits to reduce the chances of specimen or equipment damage.
58 | MTS TestLine Test System Operation
Running the Example HCF Test