2. Set up a force waveform using the Function Generator.

This introduces a force waveform to the specimen.

a) In the Station Manager main window, click the Function Generator button.
b) Set up a force waveform using the following parameters:

Channel Ch1 (or other depending on your configuration)

Control Mode Displacement

Command Type Cyclic

Target Setpoint 0.000 mm

Amplitude 2.000 mm

Frequency 1 Hz

Wave Shape Ramp

Compensator None

3. Open a Station Manager Scope.

Click the Scope button in the Station Manager main window.

4. Set up the Scope.

MTS TestLine Test System Operation | 67

Running the Example HCF Test