Use the following parameters:

Meter Type Timed

Channel Ch1 (or other depending on your configuration)

Signal Force

Engineering Units kN

Zero Displacement and Force Signals

With the actuator at mid-travel and no force on the load cell, zero the displacement and force signals.

1. Auto offset the Displacement signal.

a) Click the Station Setup button.
b) Select Channels > Ch1 > Displacement > Channel Input Signals button > Offset/Zero tab.
c) Click the Auto Offset button.

2. Auto offset the Force signal.

80 | MTS TestLine Test System Operation

Running the Example HCF Test