Retract the actuator fully, and then measure the length of the actuator so that you can later determine
where to set your limit detectors.
3. Position the actuator at the mid-point of its travel.
Calculate the mid-point. Move the actuator to the calculated mid-point to place the actuator at mid-travel.
4. Click Auto Offset so that mid-travel represents 0 mm displacement.
Turn Off Hydraulic Power
1. Turn off the HSM.
76 | MTS TestLine Test System Operation