Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

10

8/1/14

When the Program Phase number is displayed and the current Phase’s function is a rate function, a momentary
press of the ‘Program Phase #’ key exits ‘Program Entry Mode, and displays the pumping rate.

To change the current Program Phase number, press the arrow keys below the Phase number’s digits. The
maximum Phase number is 41. To reset to Phase number 1, press and hold the ‘Program Phase #’ key until the
Phase number is 1.

When a new Program Phase number is selected, the current value of all settings will be that of the currently
selected Program Phase.

5.11 ‘Program Function’ Key

When in the “Program Entry Mode”, momentary presses of the ’Program Phase #’ and the ‘Program Function’
keys switch between the Program Phase number and the Program Function displays.

With the Program Phase function displayed, the Program Function can be selected. Pressing any arrow key, or
an arrow key to the left of the colon (:) or decimal point (.) if displayed with the function, will select the next
Program Function. The selected function is stored by either pressing any non-arrow key, or after a 2 second
delay. If the selected function is different from the original function, the display will blink when the selected
function is stored.

5.11.1

Program Phase Function Parameter

If the selected function has a parameter associated with the function, the value of the parameter will be
displayed to the right of the function name, separated by either a period (.) or a colon (:).

To change the parameter’s value, press the arrow keys below the parameter’s digits. The parameter’s new
value is stored by either pressing any non-arrow key or after a 2 second delay. If the parameter has changed
from its original value, the display will blink when the parameter’s new value is stored.

5.12 ‘Setup’ Key

The secondary function of the ‘Diameter’ key is ‘Setup’. While the Pumping Program is not operating, press
and hold the ‘Diameter’ key until the first setup configuration parameter, “Power Failure Mode”, is displayed:

n

.

The display will consecutively display, for about 2 seconds, each Setup Configuration parameter and its
current setting. Pressing any non-arrow key will immediately advance to the next Setup Configuration
parameter.

To change a Setup Configuration parameter, press an arrow key under the parameter’s value. To store the new
value, press any non-arrow key or wait 2 seconds. If the parameter value differs from its previous value, the
display will blink. The new parameter value will be stored and the next parameter will be displayed. See
section 7, “Setup Configuration” for a complete description of the Setup Configurations.

After the last configuration parameter is displayed, the display reverts back to displaying the syringe diameter.
Any new parameter value will take effect immediately upon being stored.

5.13 Special Power-Up Functions

The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.

5.13.1

Firmware Version Display

To display the pump’s firmware version, press the left-most up-arrow key (

/ ) while turning on power to

the pump. The display will show:

n.nn

, where ‘n.nn’ is the firmware version number. Pressing any key

will clear the display.

5.13.2

Reset Pumping Program

To clear out the current Program Function setups, press the right-most up-arrow key (

) while turning on

power to the pump. The display will show

. Pressing any key will clear the display.

With pumps having as many complex features as the NE-1000 family of pumps, it is easy for a novice user
experimenting with the pump's setup to get the pump into a 'weird' state. Performing this reset function will
bring the pump out of a 'weird' state.