Pump reset, Quick start instructions, Setup pumping parameters – New Era NE-1600 Multi-Phaser Programmable Syringe User Manual

Page 2: Load the syringes, Prime / purge, Start the pump, When pumping

Advertising
background image

New Era Pump Systems Inc.

WWW.SYRINGEPUMP.COM

Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

i

8/1/14

Quick Start Instructions

Assumes that the pump was not previous programmed with a multiple Phase Pumping Program.

Plug in the pump.

Press the power switch to turn on power.

Press any key to stop the display from blinking.

Setup Pumping Parameters

To Change Numbers:

Use the up-arrow keys to increment individual digits.

To set/clear the decimal point

: Simultaneously press the 2 up-arrow keys under the 2 digits next to the decimal

point position. Alternatively, press and hold the left-most up-arrow key for at least 1 second. When the digit
increments from 9 to 0, the decimal point will begin to shift. Release the key when the decimal point is correct.

Press any non-arrow key, or wait 2 seconds, to enter the new setting. The display will blink when a new value is
entered and stored in memory.

Set the Syringe Inside Diameter:

Momentarily press the ‘Diameter’ key. Set the diameter.

Set the Pumping Rate.

Momentarily press the ‘Rate’ key.

To change the

pumping rate units

:

Momentarily press the ‘Rate’ key again. The display will show:

Press any up-arrow key to select the next available rate units while the units LEDs are blinking

.Press any non-arrow key, or wait 2 seconds, to set the rate units.

Set the pumping rate. If the pumping rate is out of range, the display will show:

Set the Volume to be Dispensed or Continuous Pumping

Momentarily press the ‘Volume’ key.

When the display shows:

, the pump is set for continuous pumping. Pressing any up arrow key will change

the display to 0.

For continuous pumping: Set the volume to 0.

For a Volume to be Dispensed: Set the volume. The default units are set according to the syringe diameter.

Set the Pumping Direction

When the ‘Withdraw’ LED is lit, the pump is set for withdrawing. When not lit, the pump is set for infusing. Use the

’ key to change the pumping direction.

Load the Syringes

Press in the white drive-nut button to move the pusher bar.

Insert the syringe plungers in the pusher bar bracket slots. Space unused syringe slots evenly around center.

Insert the syringe barrel flange in the flange bracket slots with the syringe barrel holder lowered onto the syringes.
Tighten the flange bracket onto the syringe flanges.

Prime / Purge:

Press and hold the ‘Start/Stop’ key for one second. Release to stop.

Start the Pump:

Press and release the ‘Start/Stop’ key to start or stop the pump.

When Pumping

The pumping rate can be changed.

With continuous pumping, the pumping direction can be changed.

PUMP RESET:

PUMP RESET:

PUMP RESET:

PUMP RESET:

Press and hold the right-most up-arrow key while turning on power to the pump.