New Era NE-1600 Multi-Phaser Programmable Syringe User Manual

Page 45

Advertising
background image

New Era Pump Systems Inc.

WWW.SYRINGEPUMP.COM

Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

40

8/1/14

Query response: { 0 | 1 }

0 = Buzzer off
1 = Buzzer is on continuously or beeping.

9.4.5

System Commands

SET PUMP NETWORK ADDRESS AND BAUD RATE

* ADR [ <address> [ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ]
[DUAL | RECP | ALTR]

Set/query pump network address

<address> => <n> [<n>]
<address> Valid range: 0 to 99

B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command

response and all further communications will be at the specified baud rate.

Special communications modes

DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.
ALTR Sets Alternating pumping mode.

All special communications modes use 19,200 baud rate.

This is a special system command that will be accepted by the pump regardless of its current address or mode.
Once set, the pump will only respond to commands with the set address and at the specified baud rate.

Note: Once a special communications mode is set, the pump will only respond to commands that are preceded
by the ‘*’ character. To exit special communication mode, reset the address: * ADR 0

Example: *ADR

Query current address setting

*ADR 3

Set pump network address to 3. The pump will now only respond to
commands with address 3

*ADR 5 B 1200

Set the pump network address to 5 and the baud rate to 1200. The
command response will be at 1200 baud

ENABLE SAFE COMMUNICATIONS MODE
SAF [ <time out> ]

Set/query Safe communications mode setting.
<time out> => <n> [ <n> [ <n> ] ]
<time out> Valid range: 0 to 255.

If <time out> = 0 then Basic communication mode is set, disabling Safe mode,

If <time out> > 0 then Safe communications mode is enabled. After the reception of this command,
valid communications must be received every <time out> seconds.

FIRMWARE VERSION QUERY
VER

Response: NE100MV <n> . <nnn>

Where ‘<n>.<nnn>‘ is current firmware version number.

Set is not applicable.

MASTER PROGRAM RESET
* RESET

Clears program memory and resets communication parameters to Basic mode and address 0.

This is a special system command that will be accepted by the pump regardless of its current address.