5 specifications, 6 custom applications – New Era NE-1600 Multi-Phaser Programmable Syringe User Manual

Page 53

Advertising
background image

New Era Pump Systems Inc.

WWW.SYRINGEPUMP.COM

Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

48

8/1/14

Using a lower baud rate may also improve the reliability of the RS-232 communications.

Pusher block makes a snap or click sound when the pump is started: This is a normal condition. When
the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive screw. The
sound heard is the drive-nut engaging on the drive screw.

Pump doesn’t stop after dispensing a set volume: The pump was previously setup with a multiple Phase
Pumping Program. To simply dispense a fixed volume at a fixed pumping rate, the second Program Phase
must be the ‘Stop’ function. See section 8.1, How to Enter Pumping Programs, for instructions on changing
the Pumping Program.

Pump stops pumping after a period of time: A dispense target volume has been set. Verify that the
dispense target volume is set to 0.

11.5 Specifications

11.5.1

Mechanical & Electrical

Model

Number of

Syringes

Maximum

Syringe Size (mL)

NE-1600

6

140

NE-1800

8

10

NE-1200

12

3


Motor type:

Step motor

Motor steps per revolution:

200

Drive screw pitch:

24 revolutions/”

Motor to drive screw ratio:

5/1

Microstepping:

1/8 to 1/2 depending on motor speed

Advance per microstep:

0.132291667

µm to 0.529166667 µm depending on motor speed


Power connector:

2.5 mm, center positive, DC

Voltage at DC connector:

12V DC at full load

Amperage:

1000 mA at full load

Power supply type:

Unregulated linear external wall adapter, country and power
source specific (or compatible regulated power supply)

Power supply output rating:

12V DC @ 1000 mA

Dimensions:

10 1/4” x 15” x 5” (LxWxH)
(26.035 cm x 38.1 cm x 12.7 cm)

Weight:

10.125 lbs. (4.595 kg)

11.5.2

Operational

Maximum speed:

3.4917 cm/min

Minimum speed:

0.0026 cm/hr

Maximum pumping rate:

1163 mL/hr with a B-D 60 cc syringe

Minimum pumping rate:

0.454

µL/hr with a B-D 1 cc syringe

Force applied to all syringes:

160 lbs. at minimum speed, 30 lbs. at maximum speed

Number of Program Phases:

41

RS-232 pump network:

100 pumps maximum

RS-232 selectable baud rates:

300, 1200, 2400, 9600, 19200

Syringe inside diameter range:

0.100 to 50.00 mm

11.6 Custom Applications

For specialized and OEM applications, contact your dealer or
New Era Pump Systems Inc. Custom modifications can be
made to the mechanics or the firmware.

Advertising