3 example 3: ramping the flow rate – New Era NE-1600 Multi-Phaser Programmable Syringe User Manual

Page 30

Advertising
background image

New Era Pump Systems Inc.

WWW.SYRINGEPUMP.COM

Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

25

8/1/14

8.4.3

Example 3: Ramping the Flow Rate

m

l/

h

r

R

a

te

Time

0

150

200

250

Phases
2 - 4

Phases
5 - 8

Phases
9-11

Note: Graph is representative only.
Pumping rate increments and decrements in steps.

Continuously ramp up and down the pumping rate. Starting at 200 mL/hr, the pumping rate will increment to
250 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Then the pumping rate will decrement to
150 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Finally, the pumping rate is incremented
back to 200 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed, then the process is repeated.

Phase

Function

Rate

Volume

Direction

1

RATE

200 mL/hr

0.1 mL

Infuse

Phase

Function

2

LP:ST

Phase

Function

Rate

Volume

Direction

3

INCR

1.0

0.1 mL

Infuse


Phase

Function

4

LP:50

Phase

Function

5

LP:ST

Phase

Function

Rate

Volume

Direction

6

DECR

1.0

0.1 mL

Infuse

Phase

Function

7

LP:99

Phase

Function

Rate

Volume

Direction

8

DECR

1.0

0.1 mL

Infuse

Phase

Function

9

LP:ST

Phase

Function

Rate

Volume

Direction

10

INCR

1.0

0.1 mL

Infuse

Phase

Function

11

LP:50

Phase

Function

12

JP:02