8.4.7

Example 7: Sub-Programs

This example shows some of the flexibilities provided by the Program Selection functions. The Pumping
Program starts by refilling the syringe with 50 mL at a fast pumping rate (Phase 1), then the Pumping Program
pauses for user sub-program selection (Phase 3). Then performs 5 dispenses of 10 mL at the selected rates,
then refills the syringe again and pauses for the next user sub-program selection.

The user is given the option of choosing one of three defined sub-programs.

1: Dispense 10 mL at 100 mL/hr (Phase 4)
2: Dispense 10 mL at 500 mL/hr (Phase 7)
3: Dispense 10 mL at 750 mL/hr (Phase 10)

After selecting the sub-program and pressing 'Start', the Pumping program continues execution at the selected
sub-program. After the 10 mL dispense, each sub-program jumps or continues with the loop counter function
(Phase 12). The first 4 loops continues Program Execution with the next user sub-program selection. After
the 5th loop, the program continues with Phase 13, which jumps back to the syringe refill function and starts
the whole program over.

Phase

Function

Rate

Volume

Direction

1

RATE

1500.0 mL/hr 50 mL

Withdraw

Phase

Function

2

LP:ST

Phase

Function

3

PR:IN

Phase

Function

4

PR:01

Phase

Function

Rate

Volume

Direction

5

RATE

100.0 mL/hr

10 mL

Infuse

Phase

Function

6

JP:12

Phase

Function

7

PR:02

Phase

Function

Rate

Volume

Direction

8

RATE

500.0 mL/hr

10 mL

Infuse

Phase

Function

9

JP:12

Phase

Function

10

PR:03

Phase

Function

Rate

Volume

Direction

11

RATE

750.0 mL/hr

10 mL

Infuse

Phase

Function

12

LP:05

Phase

Function

13

JP:01