SC-01 Operator’s Manual

Start Up • 10

Start Up

This section explains how to start up the SC-01 and to get it ready for
operation.

Start-up Procedure

1. Press and briefly hold the POWER button until you hear a beep to turn

on the SC-01.

All the alarm LEDs turn on for a few seconds, then the vibrator activates
briefly. While this is happening, all elements of the LCD display and the
LCD backlight turn on for several seconds.

2. If the CAL. LIMIT menu item in Setup Mode is on, then one of the two

following startup sequences takes place. If CAL. LIMIT is turned off, the
startup sequence in Step 3 below takes place. See “Updating the
Calibration Frequency” on page 30 for a description of the CAL. LIMIT.

If calibration is past due, the following screen appears with “CAL.”
flashing and the LEDs and buzzer pulsing several times alerting you
that the SC-01 is due for calibration. If this screen appears, wait for
the buzzer and LED pulsing to stop, then press and release the
POWER button to continue in the startup sequence.

The SC-01 will then show the current date, current battery voltage,
detection range full scale, and the alarm point settings before
proceeding to Measuring Mode and beeping twice.

If the calibration is up to date, the SC-01 then displays the
calibration due date, how many days until the next calibration,
current date, current battery voltage, detection range full scale, and
the alarm point settings before proceeding to Measuring Mode and
beeping twice.

3. If the CAL. LIMIT menu item in Setup Mode is turned off, the SC-01

shows “WARMUP” in the upper left of the LCD for a few seconds, the
displays the current date, current battery voltage, detection range full
scale, and the alarm point settings before proceeding to Measuring
Mode and beeping twice.

CAL. LIMIT

CAL.