Setup mode, Date – RKI Instruments SC-01 User Manual

Page 27

Advertising
background image

SC-01 Operator’s Manual

Setup Mode • 22

Setup Mode

This section describes the SC-01 in Setup Mode. In Setup Mode, you can
move through a menu of screens to do the following:

Set the date and time

Perform a fresh air adjustment (part of a calibration)

Perform a span adjustment (part of a calibration)

Update the alarm point settings

Update the calibration frequency

Using Setup Mode

WARNING: The SC-01 is not in operation as a gas detector while in

Setup Mode. Although it will respond to gas in parts of
AIR CAL and SPAN CAL, there are no alarm indications.

1. Take the SC-01 to a non-hazardous area and turn it off if it is on.

2. Press and hold the AIR

, then press and hold the POWER button.

When you hear a beep release the buttons. After a few seconds, the
first menu item in Setup Mode displays, the DATE Screen.

3. Use the AIR

button to move forward through Setup Mode and the

DISP button to move back through Setup Mode.

4. When you arrive at the item you wish to enter, press and release the

POWER button to enter that item.

The Setup Mode menu items are described below in the order in which
they appear while moving forward through Setup Mode.

Setting the Date and Time

Entering the DATE menu item allows you to set the date and time.

DATE

DATE

Advertising