SC-01 Operator’s Manual

Setup Mode • 30

Updating the Calibration Frequency

Entering the CAL. LIMIT menu item allows you to change the calibration
frequency.

The calibration frequency defines how often a calibration is required. It is
set in days. For example, if the calibration frequency is set to 90 days,
calibration will be due 90 days after the last successful calibration.

1. When the CAL. LIMIT Screen is displayed, press and release the

POWER button. The calibration frequency value appears and is
flashing.

2. Use the AIR

and

DISP buttons to adjust the calibration frequency

to the desired number of days. The calibration frequency is adjustable
from 1 day to 365 days. To turn the calibration frequency feature off,
adjust the frequency down to one day and continue to adjust it down
with the AIR

button until “OFF” is displayed.

3. Press and release the POWER button to accept the displayed number

of days. The unit will display the following screen before returning to the
CAL. LIMIT screen.

Entering Measuring Mode

The last screen in the Setup Mode menu is the START Screen.

When the START Screen is displayed, press and release the POWER
button to exit Setup Mode and begin the SC-01’s startup sequence in
Measuring Mode.

CAL. LIMIT

CAL. LIMIT

90

DAYS

CAL. LIMIT

End

START