SC-01 Operator’s Manual

Operation in Measuring Mode • 20

12. With the extender cable plug gently touching sensor socket, turn the

extender cable plug clockwise or counterclockwise while pushing it very
lightly until you feel it drop into place as the plug aligns with the sensor
socket and partially engages it. With the extender cable plug and
sensor socket partially engaged, the plug will no longer rotate freely.

CAUTION: Do not force the extender cable plug into the sensor socket.

Forcing a sensor into the sensor socket may damage the
sensor or the socket.

13. With the extender cable plug and sensor socket partially engaged, push

the plug into the SC-01 so that the plug fully engages the sensor
socket. This will push the plug in about 1/16” farther than the partially
engaged position.

14. Align the yellow line on the cable retainer’s rubber guard with the

“UNLOCK” line on the case and slide the cable retainer into the sensor
opening over the extender cable plug.

15. Make sure the tab on the cable retainer below the yellow line engages

the mating notch in the case’s sensor opening, then push the cable
retainer as far as it will go and turn it 1/8 turn clockwise until it locks in
place.

16. The SC-01 is now ready for use with the extender cable.