27 • Setup Mode

SC-01 Operator’s Manual

11. If the span adjustment is successful, the LCD will show the following

screen before returning to the SPAN CAL screen.

12. If the span adjustment fails, the LCD will show the following screen,

flash the alarm LEDs, and pulse the buzzer and vibrator.

When this screen appears, press and release the POWER button. The
LCD will show the following screen before returning to the SPAN CAL
screen.

13. Turn the regulator on/off knob clockwise to close it.

14. Remove the calibration cup from the sensor retainer.

15. Remove the regulator from the calibration gas cylinder.

16. Leave the regulator connected to the calibration cup for convenience.

17. Store the components of the calibration kit in a safe and convenient

place.

Updating the Alarm Point Settings

Entering the ALARM-P menu item allows you to update the SC-01’s alarm
point settings.

SPAN CAL

End

SPAN CAL

FAIL

SPAN CAL

End

ALARM-P