Fail – RKI Instruments SC-01 User Manual
Page 32

27 • Setup Mode
SC-01 Operator’s Manual
11. If the span adjustment is successful, the LCD will show the following
screen before returning to the SPAN CAL screen.
12. If the span adjustment fails, the LCD will show the following screen,
flash the alarm LEDs, and pulse the buzzer and vibrator.
When this screen appears, press and release the POWER button. The
LCD will show the following screen before returning to the SPAN CAL
screen.
13. Turn the regulator on/off knob clockwise to close it.
14. Remove the calibration cup from the sensor retainer.
15. Remove the regulator from the calibration gas cylinder.
16. Leave the regulator connected to the calibration cup for convenience.
17. Store the components of the calibration kit in a safe and convenient
place.
Updating the Alarm Point Settings
Entering the ALARM-P menu item allows you to update the SC-01’s alarm
point settings.
SPAN CAL
End
SPAN CAL
FAIL
SPAN CAL
End
ALARM-P