15 • Operation in Measuring Mode

SC-01 Operator’s Manual

Table 3: Alarm Types and Indications

Alarm Type

LCD Indications

Other Indications

Warning

Concentration of gas
rises above the
Warning level.

• Gas reading

flashes.

• Back light turns on.

• Pulsing tone

alternates with
alarm LEDs
flashing once per
second.

• Unit vibrates once

per second.

Alarm

Concentration of gas
rises above the Alarm
level.

• Gas reading

flashes.

• Back light turns on.

• Pulsing tone

alternates with
alarm LEDs
flashing twice per
second.

• Unit vibrates twice

per second.

TWA or STEL

Concentration of
target gas rises above
the TWA or STEL
alarm point.

• Back light turns on.
• Gas reading

flashes.

TWA or STEL

replaces the gas
name in the upper
left of the LCD.

• Pulsing tone

alternates with
alarm LEDs
flashing once per
second.

• Unit vibrates once

per second.

Over Range
Concentration of gas
rises above the full
scale of the SC-01.
(Or there could be a
problem with the unit.)

• Gas reading

replaced by blinking
brackets.

• Back light turns on.
• Gas name in upper

right of display
replaced by OVER.

• Pulsing tone

alternates with
alarm LEDs
flashing twice per
second.

• Unit vibrates twice

per second.

Low Battery Warning

Last remaining bar on
the right in battery icon
flashes.

None

Dead Battery Alarm

• Gas name and

reading replaced by
BATTERY FAIL.

• Battery icon

flashes.

Quick pulsing tone
and LED flashing for a
few seconds
alternates with the unit
vibrating for one
second.