25 • Setup Mode

SC-01 Operator’s Manual

WARNING: Calibrate the SC-01 in a non-hazardous environment.

1. Before performing a span adjustment, perform a fresh air adjustment as

described above in “Performing a Fresh Air Adjustment”.

2. Use the sample tubing to connect the calibration cup to the regulator.

The 81-1139RK calibration cup (used for all gases except Cl

2

and NH

3

)

has a label on it that indicates the required flow direction through the
cup. If the 81-1139RK calibration cup is used, make sure to connect the
sample tubing to the appropriate fitting so that the calibration gas flows
through the calibration cup in the direction indicated on the label.

WARNING: If the 81-1139RK calibration cup is used and the sample

tubing is not connected to the correct fitting on the
calibration cup, the calibration will not be accurate.

Table 5: Required Calibration Cups & Calibration Flow rates

Target Gas

Required

Calibration Cup

Required

Regulator,

Flowrate

Flow Direction

Specified on

Calibration Cup

Ammonia (NH

3

)

81-1139RK-NH3

81-1051RK-25,
0.25 LPM

No

Arsine (AsH

3

)

81-1139RK

81-1051RK,
0.5 LPM

Yes

Carbon Dioxide (CO)

81-1139RK

81-1051RK,
0.5 LPM

Yes

Chlorine (CL

2

)

81-1139RK-CL2

81-1051RK,
0.5 LPM

No

Hydrogen Cyanide (HCN)

81-1139RK

81-1051RK,
0.5 LPM

Yes

Hydrogen Sulfide (H

2

S)

81-1139RK

81-1051RK,
0.5 LPM

Yes

Phosphine (PH

3

)

81-1139RK

81-1051RK,
0.5 LPM

Yes

Sulphur Dioxide (SO

2

)

81-1139RK

81-1051RK,
0.5 LPM

Yes