Sensor retainer, Figure 9: installing the sensor retainer – RKI Instruments SC-01 User Manual

Page 41

Advertising
background image

SC-01 Operator’s Manual

Maintenance • 36

CAUTION: Do not force the sensor into the sensor socket. Forcing a

sensor into the sensor socket may damage the sensor or the
socket.

7. With the sensor plug and sensor socket partially engaged, push the

sensor into the SC-01 so that the sensor plug fully engages the sensor
socket. This will push the sensor in about 1/16” farther than the partially
engaged position.

8. Align the yellow line on the sensor retainer’s rubber guard with the

“UNLOCK” line on the case and slide the sensor retainer over the
sensor and push it into the sensor opening over the sensor.

9. Make sure the tab on the sensor retainer below the yellow line engages

the mating notch in the case’s sensor opening, then push the sensor
retainer as far as it will go and turn it 1/8 turn clockwise until it locks in
place.

10. Allow 5 minutes for the new sensor to stabilize.