29 • Setup Mode

SC-01 Operator’s Manual

setpoint. The unit will proceed to the TWA Screen. The TWA setpoint is
flashing.

8. Use the AIR

and

DISP buttons to adjust the TWA setpoint to the

desired value. To turn the TWA alarm off, adjust the alarm point down to
towards zero and continue to adjust it down with the AIR

button until

“OFF” is displayed.

9. Press and release the POWER button to accept the displayed TWA

setpoint. The following LCD will show the following screen before
returning to the ALARM-P screen.

The standard alarm point settings for the various models of the SC-01 are
listed below in Table 6.

Table 6: Standard Alarm Point Settings

Target Gas

Warning

Alarm

STEL

TWA

Ammonia (NH

3

)

25.0 ppm

35.0 ppm

35.0 ppm

25.0 ppm

Arsine (AsH

3

)

0.50 ppm

1.00 ppm

OFF

OFF

Chlorine (CL

2

)

0.50 ppm

1.00 ppm

1.00 ppm

0.50 ppm

Carbon Monoxide (CO)

25.0 ppm

50.0 ppm

OFF

25.0 ppm

Hydrogen Cyanide (HCN)

5.0 ppm

10.0 ppm

4.7 ppm

OFF

Hydrogen Sulfide (H

2

S)

10.0 ppm

20.0 ppm

15.0 ppm

10.0 ppm

Phosphine (PH

3

)

0.30 ppm

0.60 ppm

1.00 ppm

0.30 ppm

Sulphur Dioxide (SO

2

)

2.00 ppm

5.00 ppm

5.00 ppm

2.00 ppm

TWA

5.00

PPM

ALARM-P

End