SC-01 Operator’s Manual

Setup Mode • 24

3. Press and hold the AIR

▼

button. The LCD displays “hold” prompting

you to hold the AIR

▼

button.

4. After a moment, the LCD will indicate “Adj.”. Continue to hold the AIR

▼

button.

5. Release the AIR

▼

button when the LCD displays “RELEASE” in the

upper left corner.

6. The unit will take a few seconds to set the reading to 0 ppm, then

indicate “FINISH” in the upper left corner of the LCD and return to the
AIR CAL Screen.

Performing a Span Adjustment

Entering the SPAN CAL menu item allows you to perform a span
adjustment.

Perform a span adjustment as part of a calibration after performing a fresh
air adjustment. Performing a span adjustment requires the use of a
calibration kit. The procedure below describes a span adjustment using a
calibration kit that includes a calibration gas cylinder, a 0.5 LPLM (liters per
minute) or 0.25 LPM fixed flow regulator with an on/off knob, one of three
different calibration cups, and non-absorbent sample tubing. The required
calibration cup and calibration flow rate will depend on the type of SC-01.
The direction of flow through the cup may also be specified depending on
the cup. See Table 5 below for the calibration cup. Calibration flow rate and
required regulator, and whether the calibration cup flow direction is
specified for the various versions of the SC-01. See “Parts List” on page 38
for available calibration kits and calibration kit spare parts.

AIR CAL

h o l d

AIR CAL

Adj.

RELEASE

Adj.

SPAN CAL