Specifications • 2

Specifications

Table 1: 01 Series Specifications

Available Target Gas/
Detection Range/
Display Increment

• Ammonia (NH

3

): 0 - 75.0 ppm, 0.5 ppm increments

• Arsine (AsH

3

): 0 - 1.50 ppm, 0.01 increments

• Carbon Monoxide (CO): 0 - 75.0 ppm, 0.5 ppm increments
• Chlorine (Cl

2

): 0 - 3.00 ppm, 0.02 ppm increments

• Hydrogen Cyanide (HCN): 0 - 15.0 ppm, 0.1 increments
• Hydrogen Sulfide (H

2

S): 0 - 30.0 ppm, 0.2 ppm increments

• Phosphine (PH

3

): 0 - 1.00 ppm, 0.01 ppm increments

• Sulphur Dioxide (SO

2

): 0 - 6.00 ppm, 0.05 ppm increments

Detection Principle

Electrochemical

Sampling Method

Diffusion

Response Time

90% in 60 seconds

Accuracy

±10% of reading or ± 5% of detection range full scale
(whichever is greater)

Indication

Digital LCD

Safety/Regulatory

C US

186718

CSA classified, “C/US”, as Intrinsically Safe. Exia. Class I,

Groups A, B, C, & D. Temperature Code T3.

Power Source

Two AA size Alkaline Batteries Standard, Duracell MN1500 or
PC1500

Continuous
Operating Hours

At 25 °C: Over 250 Hours, No Alarms or Backlighing

Case

High-impact Plastic, Dust and Weather Proof

Standard
Accessories

• Alligator Clip
• Rubber protective boot

Optional Accessories

• Calibration Cup
• Calibration Kit
• Belt Clip
• 10 ft. Extender Cable
• Data Logging Kit

Dimensions and
Weight

• 2.5” W x 5.2” H x 1.2” D (63 mm x 131 mm x 31 mm)
• 7.6 ounces (215 g)