23 • Setup Mode

SC-01 Operator’s Manual

1. When the DATE Screen is displayed, press and release the POWER

button. A screen appears with the year flashing in the upper left, the
date in the middle, and the time in the lower right.

2. Use the AIR

▼

or the

▲

DISP button to display the desired year.

3. Press and release the POWER button to save the setting. The month

setting flashes.

4. Repeat steps 2 and 3 to enter the month, day, hour and minute setting.

When you save the minute setting, you return to the DATE Screen.

Performing a Fresh Air Adjustment

Entering the AIR CAL menu item allows you to perform a fresh air
adjustment.

Perform a fresh air adjustment when you are performing a calibration
before proceeding to the next menu item, SPAN CAL, to perform a span
adjustment.

WARNING: Calibrate the SC-01 in a non-hazardous environment.

1. Find a fresh air environment, an environment of normal oxygen content

(20.9%) that is free of toxic or combustible gases.

2. When the AIR CAL screen is displayed, press and release the POWER

button. A screen appears that displays the current gas reading and
prompts you to press the AIR

▼

button with a flashing “PUSH AIR”

message at the top of the screen.

2006

12.11

08:00

AIR CAL

PUSH AIR

0.00

PPM