Preparation and check before using

Caution

1.

Confirmation of outlet

„ Please insert the power supply plug in the

outlet.

„ Please confirm whether the Refrigerator plug is

connected with the outlet in the back of the
controller. (BH200/300)

2.

Confirmation of safety

„ Please turn on the leak breaker.
„

Please pour the liquid (water) into the
examination tank.

„ Water level should be 3∼5 cm below the upper

edge. The external is made to circulate and the
pump is operated and the water level falls when
passing for a while when using. Please make to
the above-mentioned water level by injecting the
liquid again.

„ Please confirm whether independent tempera-

ture overheating prevention device is set in a
temperature which is 15

℃ or more higher than

the temperature of the tank used. (BH500)

„ Please stabilize by using stopper in the bottom

in the tank when you use the throw type cooling
device together. (BH400/500)

Warning

„ To avoid the shock hazard, the use of the ground outlet is recommended.
„ Please use the earth adapter (attached goods outside the standard) when you use the outlet of the

dipoles type.

„ Please connect the earth wire of the earth adapter with the earth of the power supply equipment side.

Please do not connect to gas pipe and water pipe.

„ Please do not use the divergence outlet. Because heat etc. are caused, it is dangerous.
„ Please do not ground to the place where a flammable gas and corroded gas are generated.

Ground outlet