Preparation and check before using
Caution
1.
Confirmation of outlet
Please insert the power supply plug in the
outlet.
Please confirm whether the Refrigerator plug is
connected with the outlet in the back of the
controller. (BH200/300)
2.
Confirmation of safety
Please turn on the leak breaker.
Please pour the liquid (water) into the
examination tank.
Water level should be 3∼5 cm below the upper
edge. The external is made to circulate and the
pump is operated and the water level falls when
passing for a while when using. Please make to
the above-mentioned water level by injecting the
liquid again.
Please confirm whether independent tempera-
ture overheating prevention device is set in a
temperature which is 15
℃ or more higher than
the temperature of the tank used. (BH500)
Please stabilize by using stopper in the bottom
in the tank when you use the throw type cooling
device together. (BH400/500)
Warning
To avoid the shock hazard, the use of the ground outlet is recommended.
Please use the earth adapter (attached goods outside the standard) when you use the outlet of the
dipoles type.
Please connect the earth wire of the earth adapter with the earth of the power supply equipment side.
Please do not connect to gas pipe and water pipe.
Please do not use the divergence outlet. Because heat etc. are caused, it is dangerous.
Please do not ground to the place where a flammable gas and corroded gas are generated.
Ground outlet
Power supply plug