Specification table – Yamato Scientific BH300 Thermo-Elite User Manual

Page 32

Advertising
background image

29

Specification table

Form

BH200

BH300

Performance

↓↓↓↓

Range of temperature

setting (

℃)

-30.0∼+100

Range of operation

temperature (

℃)

-20∼80

-25∼80

Temperature adjustment

accuracy (

℃)

±0.01℃∼0.02℃ (Ethanol 50% sol. ,RT25℃,No-load)

Ability of circulating pump

Max. pump head : 3(mH

2

O), Max. stream flow : 20(l/min)

(Refrigerant : Water , Water temperature : 25℃)

Cooling ability (kcal/h)

150(at20℃,RT25℃)

200(at20℃,RT25℃)

Function and composition

↓↓↓↓

Temperature adjustment

machine

IVCR type controller (Zero crossover switching, PID microprocessor control)

Sensor

Pt100Ω A class (three line type)

Temperature set method

Digital setting depending on ▲▼keys

Temperature display method

Digital on a main display (A set temperature is displayed to a sub display)

Resolution of setting and display

temperature

0.1℃

Other displays

Drive monitor (represents the state of control)

Timer function

1 minute to 99 hours 59 minutes , 100 hours to 999 hours

Timer resolution

1 minute or 1 hour

Drive function

Fixed value drive, Auto start drive,, Auto stop drive

Program drive : Arbitrary pattern, Max.16 segment (repetition and ramp operation,

etc.) , Refrigerator drive function (There is no refrigerator energizing indicator)

Output terminal

(Seven pin

connector)

Other incidental functions

Memory backup

Key lock

Multiplication

time function

( up to 49999

hours)

Holding

Function that

enable to change

the strength of jet

stream (10 step

changeability)

Calendar timer

Clock

Analogue

External

warning output

Time-up output

External

communication

(RS422A

correspondence)

Communication

terminal

(eight pin

connectors)

Heater (made of SUS)

1.0kw

stirring mechanism

Stirring by jet stream

Compression

machine output

150W

400W

Refrigerator

Refrigerant

R22

Other compositions

Drain entrance (with stopping) , Stainless flexibility tube with heat insulation for

circulation

Leak Breaker

Thermal protector (for stirring motor)

The float type High/Low liquid level detector (Low level only for BH400)

Self-diagnosis function (Automatic overheat protector, The sensor is abnormal,

Heater disconnection, Triac short-circuit, Main relay is defective, Buzzer warning

in abnormal circumstances, The refrigerator is abnormal)

Safety device

Refrigerator automatic operation stop (It is shutdown at 45℃ or more in set

temperature)

Standard

↓↓↓↓

Capacity of reservoir ( l )

13

30

Size of reservoir

(W

×D×H mm)

389Ч226Ч150

501Ч295Ч200

The outside size

(W

×D×H mm)

654Ч492Ч615

385Ч589Ч1150

Power supply

AC100V 13A

AC100V 14A

Standard attached goods

Stainless hose entrance for external circulation (Rc1/2×Outside diameterφ14mm)×2

Digital printer

independent temperature overheating prevention device

By-pass tube for external circulation, Insulated hose for external circulation

Attached goods outside standard

Cable for external communication, Personal computer connection interface,

(RC23), Multi wick cable for output (All model commonness)