Incidental function

Usage of external output

Attention : The program method, the set method, and the communication interface are described in

detail to the appended owner guide only for the controller. Please refer to it.

1.

Specification of connector

From the back of Thermo Elite

Pin number

Name of signal

DC+12V

Time-up output(open collector)

DC+12V

External warning output (open collector)

Temperature output (+)

Temperature output (-)

Shield (analog signal)

2.

Notes concerning liquid used

z

Unit of output voltage : 5mV/℃

z

The output is 0mV at 0℃ in measurement temperature.

Output

voltage(V)

Measurement temperature(

℃)

3.

External warning output

4.

Time-out output

z

Transistor open collector output

z

Transistor open collector output

Output load

Power supply DC12V

Output load

Power supply DC12V

sink current

MAX 100mA

sink current

MAX 100mA

Specification of
out-put relay

MATSUSHITA JAI-F-TM-
DC12V / OMRON G2R-1A-T
DC12V

Specification of
out-put relay

MATSUSHITA JAI-F-TM-
DC12V / OMRON G2R-1A-T
DC12V

TROUBLE sign
(Blink to a main
display the err-or
code)

Blinking

Blink to the FI-
NISH indicator

Blinking

External warning
output

ON

Time-out output

ON